Miranda Holley, Gabriela (2009) Control de calidad de filtros compensadores discretos para IMRT. Master in Medical Physics, Universidad Nacional de Cuyo, Instituto Balseiro.
![]()
| PDF (Tesis) Spanish 1787Kb |
Abstract in Spanish
Se ha implementado radioterapia de intensidad modulada en la FUESMEN, desarrollándose un nuevo método de filtros compensadores, discretizados en capas y confeccionados en el mismo Servicio. En este trabajo se realizaron mediciones para el control de calidad de estos accesorios para varios casos: un compensador de geometría controlada, filtros resultantes de la planificación inversa en un fantoma elipsoidal y para los dos primeros pacientes. Se analizó la concordancia entre las mediciones con un arreglo de detectores de estado sólido y los cálculos del planificador, usando el índice γ con criterios de aceptación de 6%/3mm para los campos individuales y de 4%/3mm para la combinación de los campos que componen un plan, obteniéndose más de un 90% de puntos aceptados en todos los casos. Se realizó asimismo un análisis cualitativo con films radiocrómicos, evaluando la coincidencia de curvas de isodosis medidas y calculadas. También se realizaron cálculos independientes de unidades monitoras para los planes de ambos pacientes. Además se estudiaron algunos de los parámetros que utiliza el planificador computado para la optimización del plan. Se concluyó que los filtros compensadores desarrollados cumplen con los criterios de aceptación establecidos, y pueden utilizarse para tratamientos sin alteraciones significativas en las dosis. Palabras claves: IMRT, filtros compensadores, QA, MapCHECK, film radiocrómico.
Abstract in English
A new model of compensating filter for intensity-modulated radiotherapy has been developed at FUESMEN; they have been designed by discrete layers and handmade at the radiotherapy Service. In the present work, the quality control measurements for this type of accessories are performed for various cases: a geometry-controlled compensator, the filters resulting from an inverse planning in an ellipsoidal phantom and for the first two treated patients. The correspondence between the dosimetry with a 2D diode array and the planned distributions was analyzed with gamma index criteria of 6%/3mm for individual beams, and 4%/3mm for a composite plan, obtaining over 90% agreement in all studied cases. A qualitative analysis was carried out using radiochromic film dosimetry, assessing the coincidence between measured and planned isodoses. Independent calculations of monitor units were made for the patients’ plans as well. Furthermore, some parameters of the treatment planning software for plan optimization were studied. It was concluded that the developed compensating filters fulfill the established acceptance criteria, and can be used for treatment without significant dose deviation.
Item Type: | Thesis (Master in Medical Physics) |
---|---|
Keywords: | Radioterapia; Radiotherapy; Control de calidad; Quality control; Filtros; Filters; Dosimetría; Dosimetry; Garantía de calidad; Quality assurance; Films radiocrómicos |
Subjects: | Medicine |
Divisions: | FUESMEN |
ID Code: | 116 |
Deposited By: | Marisa G. Velazco Aldao |
Deposited On: | 15 Jun 2010 15:49 |
Last Modified: | 15 Jun 2010 15:49 |
Repository Staff Only: item control page